Using Collision Cross Section Distributions to Assess the Distribution of Collision Cross Section Values

20 January 2020, Version 2

Abstract

In this study we explore the use of collision cross section distributions to allow comparability of IM-MS data for proteins on different instruments. We present measurements on seven standard proteins across three IM-MS configurations, namely an Agilent 6560 IM QToF, a Waters Synapt G2 possessing a TWIMS cell and a modified Synapt G2 possessing an RF confining linear field drift cell. Mobility measurements were taken using both He and N2 as the drift gases. To aid comparability across instruments and best assess the corresponding gas-phase conformational landscapes of the protein ‘standards’ we present the data in the form of averaged collision cross section distributions.

Content

Supplementary materials

France ccs compare SI sub
France ccs compare paper version2
France ccs compare SI version2
TWIMS CCS Calibration plots

Comments

Comments are not moderated before they are posted, but they can be removed by the site moderators if they are found to be in contravention of our Commenting Policy [opens in a new tab] - please read this policy before you post. Comments should be used for scholarly discussion of the content in question. You can find more information about how to use the commenting feature here [opens in a new tab] .