Experimental Strategies for Avoiding Saturation Effects in ESI-MS

16 July 2019, Version 1
This content is a preprint and has not undergone peer review at the time of posting.

Abstract

All instruments with detectors are prone to saturation effects at high concentration, and mass spectrometers are no exception. The very high sensitivity of mass spectrometry makes the onset of saturation occur at lower concentrations than other methods, and in cases where the analyte of interest is very reactive, concentrations at which saturation can be problematic may be necessary in order to ensure decomposition is mitigated. Indications that saturation is occurring are provided, and some data processing strategies are outlined, followed by a range of detuning strategies that may be employed to reduce saturation effects in the context of electrospray ionization mass spectrometry (ESI-MS), including lowering voltages on detector or capillary, increasing cone gas flow rates, or adjusting the probe position. A combination of strategies generally allows researchers to make the best possible compromises when studying compounds at relatively high concentration.

Keywords

mass spectrometry
electrospray ionization
saturation
decomposition

Supplementary materials

Title
Description
Actions
Title
Manuscript
Description
Actions
Title
SupportingInformation
Description
Actions

Comments

Comments are not moderated before they are posted, but they can be removed by the site moderators if they are found to be in contravention of our Commenting Policy [opens in a new tab] - please read this policy before you post. Comments should be used for scholarly discussion of the content in question. You can find more information about how to use the commenting feature here [opens in a new tab] .
This site is protected by reCAPTCHA and the Google Privacy Policy [opens in a new tab] and Terms of Service [opens in a new tab] apply.